Casimir Pressure Derivation
Appendix J
· Casimir-type pressure in Phase-Biased Geometry
C 1 Problem statement
Two perfectly conducting, plane plates are held parallel in vacuum at a gap
Goal: derive from the PBG action
the pressure
Throughout we use the already-calibrated substrate constants
The associated wave-number
so all practical gaps satisfy
C 2 Mode structure between plates
Boundary conditions (perfect conductors):
Allowed standing waves
with transverse wave-vector
C 3 Anchoring cost without the time-kinetic term
(Static contribution only)
For each mode
(A) = plate area. Summing and subtracting the free-space continuum gives
and therefore
(pressure falls as
C 4 Fluctuation (zero-point) contribution – the role of γ
Restore the
Each mode carries vacuum energy
Repeating the same zeta-regularised sum (now over
identical to the standard QED Casimir formula because
C 5 Full PBG prediction
Add the two independent pieces:
Magnitude check
Gap (d) | Static term (Pa) | Fluctuation term (Pa) | Total |
---|---|---|---|
100 nm | |||
50 nm | |||
10 nm |
⇒ the static part is negligible ( less than 1 %) for
Laboratory measurements at 50–500 nm therefore test primarily the fluctuation piece (C-3).
C 6 Comparison with experiment
- Recent Au–Au torsion-balance data at 0.1–0.5 µm gaps report agreement with the
law to ±5 %. - That matches the PBG total (C-4) because the α/d³ contribution is
in this range. - A decisive PBG-only signature would be to probe
where the term reaches ≥1 % of the total; current metrology does not yet resolve forces at the kPa level for such tiny gaps.
C 7 What would falsify the model?
- Exponent test If future high-precision data confirmed a pure
law down to, say, 5 nm with no hint of a admixture at the 1 % level, the minimal PBG action would need modification (e.g. an extra counter-term that cancels (C-2)). - Material test The fluctuation term depends only on
and ; the static term depends on α but not on conductivity. Any detected material dependence in the 10-100 nm regime would contradict equation (C-4). - Amplitude test A measured prefactor differing by more than the experimental ±5 % from
would also rule out the present PBG constants because (c) is fixed and no other tunable parameter exists.
C 8 Concluding remark
With no additional constants beyond